d) In the appearing window: Check the required system → Click Next
e) In the appearing dialog box: Select the spectrometer configuration name →
Click Next
f) In the appearing dialog box: Check the items you want to install or, accept
the default selection → Click Next
g) In the appearing dialog box: Select the default printer and plotter and the
desired paper format → Click Next
h) In the appearing window: Enter the desired spectrometer frequency, acquisi-
tion mode and pre-scan-delay or, accept the default values → Click Next
i) In the appearing information box: Check the configuration selection and, if
it is correct → Click Finish
and wait ...
• Continue with edsolv
Check the list of lock solvents → click Save if it is correct → Click Close
• Continue with edhead
Select the current probe → click Define as current probe → Click Exit
A window Edit Spectrometer Parameter will appear:
Check if the connections between preamplifier and probe are set correctly →
click Save if it is correct → Click Close
• Continue with edprosol
Check the parameters for your observed nuclei, and if necessary change
and/or recalculate them → Click Save to store to disk → Click Exit
• Continue with edlock
Check the lock table values (field, lock phase, ...), change them and click
Save if necessary, otherwise click Abort
• Continue with edscon
Check the spectrometer constants, change them and click Save if necessary,
otherwise click Cancel
• After finishing the configuration a window
cron check appears. Click on
Automatic backup to set up an periodically backup of your TOPSPIN
spectrometer configuration.
If you click OK in the following window an automatic backup every 4 weeks
will be initiated.